Set up Trace B to measure velocity in inch/s (0-pk):

AGILENT 35670A

Supplemental Operator’s Guide

Section 19 : Two Spectral Traces Showing Mils and Ips while EU is G

Measure an accelerometer signal and display velocity and displacement spectra with the UPPER/LOWER display format in the AGILENT 35670A as follows:

Preset the AGILENT 35670A

Turn on the AGILENT 35670A, or

Press the green [Preset] hardkey

Press the yellow [Pause/Cont] hardkey if desired.

Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.

Set up transducer parameters for a 10 mV/g accelerometer:

Press the [Input] hardkey

Press [FRONT END CH1 SETUP] softkey (F7).

Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.

Press [Input] hardkey.

Press [XDCR UNIT CH1 SETUP] softkey (F8)

Press the [XDCR SENSITVTY] softkey (F6).

Enter 10 (for an accelerometer with 10 mV/g sensitivity), then press the

[mV/EU] softkey (F2).

Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.

Press the [XDCR UNIT LABEL] softkey (F7).

Select the [g] softkey (F2).

Set up display format and Trace A to measure displacement in mil (0-pk):

Press the [Disp Format] hardkey.

Press the [UPPER/LOWER] softkey (F3)

Press the [Active Trace] hardkey.

Press the [ A ] softkey (F5).

Press the [Trace Coord] hardkey.

Press the [Y UNITS] softkey (F8)

Toggle the [AMPLITUDE PK PP RMS] softkey (F4) until PK is highlighted.

Press the [XDCR UNIT CONVERT] softkey (F4)

Press the [mil (DISP)] softkey (F8)

Press the [Scale] hardkey.

Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.

Press the yellow [Start] hardkey.

Set up Trace B to measure velocity in inch/s (0-pk):

Press the [Active Trace] hardkey.

Press the [ B ] softkey (F2).