Section 15 : Waterfall Spectra at Time Intervals
Waterfalls on the AGILENT 35670A can be used to examine and compare spectra at fixed intervals of time. This is especially useful when measuring
Preset the analyzer as follows:
•Press the green [Preset] hardkey
•Press [DO PRESET] softkey (F1).
•Press the [Pause/Cont] hardkey if desired.
Selecting Single Channel Operation
Specify the
35670A and then:
•Press the [Inst Mode] hardkey
•Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.
Selecting Measurement Parameters
Refer to the section entitled “Measuring Single Channel Spectra” to set up span, ranges, coupling, averaging, and windowing. Transducer parameters set up is described in the section “Measuring with Transducers.”
Setting Up the Waterfall Display
The best view of a waterfall includes the current power spectrum above the waterfall trace. In the AGILENT35670A this is implemented with the UPPER/BIG LOWER display format as follows:
•Press the [Disp Format] hardkey.
•Select the [UPPER/ BIG LOWER] softkey (F4).
•Press the [Active Trace] hardkey.
•Select the [A] softkey (F2).
•Press the [Meas Data] hardkey.
• Toggle the [CHANNEL 1 2 3 4] softkey until 1 is highlighted.
•Press the [PWR SPEC CHANNEL 1] softkey (F3).
Continuing to set up a waterfall display,
•Press the [Active Trace] hardkey.
•Select the [B] softkey (F2).
39