Section 15 : Waterfall Spectra at Time Intervals

AGILENT 35670A

Supplemental Operator’s Guide

Section 15 : Waterfall Spectra at Time Intervals

Waterfalls on the AGILENT 35670A can be used to examine and compare spectra at fixed intervals of time. This is especially useful when measuring non-stationary signals from machinery, such as during run-ups and run-downs. To see a waterfall display of the vibration signature, follow these steps:

Preset the analyzer as follows:

Press the green [Preset] hardkey

Press [DO PRESET] softkey (F1).

Press the [Pause/Cont] hardkey if desired.

Selecting Single Channel Operation

Specify the single-channel instrument mode to measure one waterfall. Note that it is possible to measure and display waterfalls for two channels simultaneously. For single channel operation, however, the display performance will be better with only one channel active. To measure a single-channel waterfall, turn on the AGILENT

35670A and then:

Press the [Inst Mode] hardkey

Press [CHANNELS 1 2 4 ] softkey (F7) until 1 is highlighted.

Selecting Measurement Parameters

Refer to the section entitled “Measuring Single Channel Spectra” to set up span, ranges, coupling, averaging, and windowing. Transducer parameters set up is described in the section “Measuring with Transducers.”

Setting Up the Waterfall Display

The best view of a waterfall includes the current power spectrum above the waterfall trace. In the AGILENT35670A this is implemented with the UPPER/BIG LOWER display format as follows:

Press the [Disp Format] hardkey.

Select the [UPPER/ BIG LOWER] softkey (F4).

Press the [Active Trace] hardkey.

Select the [A] softkey (F2).

Press the [Meas Data] hardkey.

Toggle the [CHANNEL 1 2 3 4] softkey until 1 is highlighted.

Press the [PWR SPEC CHANNEL 1] softkey (F3).

Continuing to set up a waterfall display,

Press the [Active Trace] hardkey.

Select the [B] softkey (F2).

39

Page 43
Image 43
Agilent Technologies Agilent 35670A manual Setting Up the Waterfall Display, Waterfall Spectra at Time Intervals