Connect the transducers

AGILENT 35670A

Supplemental Operator’s Guide

Attach one end of stinger to shaker and one end to force transducer

Connect amplifier output to shaker

Connect amplifier input to AGILENT 35670A source

Connect the transducers

Connect force transducer to Channel 1

Connect accelerometer to Channel 2

Preset the AGILENT 35670A

Turn on the AGILENT 35670A, or

Press the green [Preset] hardkey

Press the yellow [Pause/Cont] hardkey if desired.

Enter Channel 1 input parameters

Press the [Input] hardkey

Press [CH1 AUTO UP ONLY] softkey (F5).

Press [FRONT END CH1 SETUP] softkey (F7).

Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.

Press the [Input] hardkey

Press [XDCR UNIT CH1 SETUP] softkey (F8)

Press the [XDCR SENSITVTY] softkey (F6).

Enter 8, then press the [mV/EU] softkey (F2).

Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.

Press the [XDCR UNIT LABEL] softkey (F7).

Press the [MORE CHOICES] softkey (F9).

Select the [ lb ] softkey (F6).

Enter Channel 2 input parameters

Press the [Input] hardkey

Press [CHANNEL 1 2 3 4] softkey (F1) until 2 is highlighted.

Press [CH2 AUTO UP ONLY] softkey (F5).

Press [FRONT END CH2 SETUP] softkey (F7).

Press [ICP SUPLY ON OFF] softkey (F8) until ON is highlighted.

Press the [Input] hardkey

Press [XDCR UNIT CH2 SETUP] softkey (F8)

Press the [XDCR SENSITVTY] softkey (F6).

Enter 10.2, then press the [mV/EU] softkey (F2).

Toggle the [XDCR UNIT ON OFF] softkey (F4) until ON is highlighted.

Press the [XDCR UNIT LABEL] softkey (F7).

Select the [ g ] softkey (F2).