Choose measurement parameters

Press the [Freq] hardkey

Enter 800 and then press the [Hz] softkey (F2)

Press the [Avg] hardkey.

Press the [NUMBER AVERAGES] softkey (F2), enter 4, then press the

[ENTER] softkey (F1).

Turn on averaging by pressing the [AVERAGE ON OFF] softkey (F1) until ON is highlighted.

Press the [Window] hardkey

Press the [UNIFORM] softkey (F3).

Set up display parameters

Press the [Disp Format] hardkey.

Press the [UPPER/LOWER] softkey (F3)

Press the [Active Trace] hardkey.

Press the [ AB ] softkey (F5).

Press the [Meas Data] hardkey.

Press the [FREQ RESP 2 / 1 ] softkey (F6)

Press the [Scale] hardkey.

Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.

Press the [Active Trace] hardkey.

Press the [ B ] softkey (F2).

Press the [Trace Coord] hardkey.

Press the [PHASE] softkey (F4).

Set up the AGILENT 35670A source (if amplifier has level control)

Check the following connections:

AGILENT 35670A source connected to amplifier input amplifier plugged in

amplifier output connected to shaker force transducer connected to Channel 1

Reduce amplifier level control the lowest setting

Turn on amplifier

Press the [Source] hardkey.

Press the [RANDOM NOISE] softkey (F5)

Press the [LEVEL] softkey (F2)

Enter 0.5, then press the [Vpk] softkey (F1).

Toggle the [SOURCE ON OFF] softkey (F1) until ON is highlighted.