AGILENT 35670A
Supplemental Operator’s Guide
Choose measurement parameters
•Press the [Freq] hardkey
•Enter 800 and then press the [Hz] softkey (F2)
•Press the [Avg] hardkey.
•Press the [NUMBER AVERAGES] softkey (F2), enter 4, then press the
[ENTER] softkey (F1).
•Turn on averaging by pressing the [AVERAGE ON OFF] softkey (F1) until ON is highlighted.
•Press the [Window] hardkey
•Press the [UNIFORM] softkey (F3).
Set up display parameters
•Press the [Disp Format] hardkey.
•Press the [UPPER/LOWER] softkey (F3)
•Press the [Active Trace] hardkey.
•Press the [ AB ] softkey (F5).
•Press the [Meas Data] hardkey.
• Press the [FREQ RESP 2 / 1 ] softkey (F6)
•Press the [Scale] hardkey.
•Toggle the [AUTOSCALE ON OFF] softkey (F1) until ON is highlighted.
•Press the [Active Trace] hardkey.
•Press the [ B ] softkey (F2).
•Press the [Trace Coord] hardkey.
•Press the [PHASE] softkey (F4).
Set up the AGILENT 35670A source (if amplifier has level control)
•Check the following connections:
AGILENT 35670A source connected to amplifier input amplifier plugged in
amplifier output connected to shaker force transducer connected to Channel 1
•Reduce amplifier level control the lowest setting
•Turn on amplifier
•Press the [Source] hardkey.
•Press the [RANDOM NOISE] softkey (F5)
•Press the [LEVEL] softkey (F2)
•Enter 0.5, then press the [Vpk] softkey (F1).
•Toggle the [SOURCE ON OFF] softkey (F1) until ON is highlighted.
52