Troubleshooting

Troubleshooting Gateway Configuration

4If the Gateway is powered on, configured, but you do not know the IP address:

n Press the Config Preset button on the back panel of the LAN/GPIB Gateway. This forces the Gateway to temporarily use the 192.0.0.192 default IP address without modifying any other configuration values.

n Make sure the route table entry for the Gateway at its default IP address is setup via the route command on the client system.

n Telnet to 192.0.0.192 and look at the setting of the bootp: ON/OFF configuration value to determine the configuration method used.

NOTE

If you are unable to make a Telnet connection using the default IP address (192.0.0.192), try pressing the Config Preset button again and then Telnet to 192.0.0.192.

nExecute a Telnet reboot command. If no values were changed, the Gateway will use the same configuration method and have the same configuration values (IP address, etc.) as previously set.

CAUTION

Any client operations in progress and client connections will be terminated by the reboot command.