Gateway Failure Messages

Troubleshooting

Gateway Failure Messages

Gateway Failure Messages

This section provides failure messages and troubleshooting explanations for the E2050 LAN/GPIB Gateway, including:

nSICL Error Codes and Messages

nsyslog File Messages

SICL Error Codes and Messages

The following table lists some common SICL error codes and messages a client application may encounter when using the E2050 LAN/GPIB Gateway.

Error Code

Error Message

Troubleshooting Tips

 

 

 

I_ERR_BADADDR

Bad address

The device/interface address passed to iopen does not

 

 

exist. Use the Telnet commands to verify that the hpib-

 

 

address:, hpib-name:, and hpib-unit: configuration

 

 

values are correct.

 

 

 

I_ERR_NOCONN

No connection

The communication session with the Gateway was never

 

 

established or the connection to a remote Gateway was

 

 

dropped. Use the Telnet status command and any error

 

 

messages logged to a syslog file for more information.

 

 

 

I_ERR_NORSRC

Out of resources

No more resources are available. Use the Telnet status

 

 

command and any error messages logged to a syslog file

 

 

for more information.

 

 

 

I_ERR_NOTSUPP

Operation not

May occur when using VXI-11 (TCP/IP Instrument Protocol)

 

supported

when the hpib-nameis set to hpib. Use the Telnet

 

 

configuration utility to change the hpib-nameto gpib0.

 

 

If this does not solve the problem, use the SICL LAN protocol

 

 

instead, as this protocol has more features than VXI-11.

 

 

 

I_ERR_SYMNAME

Invalid symbolic

The symbolic name passed to iopen was not recognized.

 

name

Use the Telnet hpib-name:configuration value to verify

 

 

the configuration.

 

 

 

I_ERR_TIMEOUT

Timeout occurred

A timeout occurred while performing an operation. The

 

 

device may be busy, in a bad state, or a longer timeout value

 

 

may be needed for the operation on the device. Verify the

 

 

correct address was used in iopen. Use the Telnet status

 

 

command and any error messages logged to a syslog file

 

 

to get more information.

 

 

 