Administration

Configuring the Gateway

To change the configuration method used for the Gateway, you can modify the bootp: ON/OFF Telnet configuration value or you can press the Config Preset button. Once you have changed the configuration method, follow the procedures in Chapter 3 to configure the Gateway via the new method.

NOTE

The bootp: ON/OFF configuration value in the Telnet utility is automatically set to OFF if the IP address is changed via the Telnet utility.

To modify the bootp: ON/OFF Telnet configuration value:

1To change to the Telnet configuration method, at the Telnet prompt (>), type: bootp: OFF

2To change to BOOTP or BOOTP with TFTP configuration method, at the Telnet prompt (>), type bootp: ON

3After you have modified the bootp: ON/OFF value, execute the reboot Telnet command to save the new configuration setting and reboot the Gateway.

CAUTION

Any client operations in progress and client connections will be terminated by the reboot command when it reboots the Gateway.

4Execute a Telnet factory-resetcommand. This changes the configuration values to the default values and reboots the Gateway. This will set the bootp: ON/OFF value to the default ON value. You may now use the BOOTP or the BOOTP with TFTP configuration method.

CAUTION

Any client operations in progress and client connections will be terminated by the factory-resetcommand when it reboots the Gateway.