Agilent Technologies U2761A manual Soft Front Panel Operation, Remote Interface Operation

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 107
Image 107

Features and Functions

2

When the Internal (immediate) trigger source is selected, the U2761A outputs a Square waveform with a 50% duty cycle to the Trig Out connector at the beginning of the sweep. The period of the Square waveform is equal to the specified sweep time.

When the Manual (software) trigger source is selected, the U2761A outputs a pulse to the Trig Out connector at the beginning of each sweep.

Soft Front Panel Operation

After enabling sweep, to specify whether the U2761A outputs a TTL-compatible Square waveform with either a rising or

falling edge, select Positive falling edge, select Positive  for rising edge and for rising edge and

Negative Manual background for falling edge.

Remote Interface Operation

OUTPut:TRIGger {OFFON}

OUTPut:TRIGger:SLOPe {POSitiveNEGative}

U2761A User’s Guide

89

Page 107
Image 107
Agilent Technologies manual Soft Front Panel Operation, Remote Interface Operation, U2761A User’s Guide