Agilent Technologies U2761A manual Set Output Using SCPI Commands, Remote Interface Operation

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 60
Image 60

2 Features and Functions

Remote Interface Operation

OUTPut {OFFON}

The APPLy command overrides the current setting and automatically enables the output control.

Set Output Using SCPI Commands

The following SCPI commands show a sample procedure of generating output.

Example 1, To output a DC voltage

-> *CLS; *RST

//To reset the U2761A to default

 

power-on state, this command

 

can be ignored if this operation

 

is not required.

-> FUNC DC

//Sets the output to DC.

-> VOLT:OFFS 1

//Sets the output DC offset voltage

 

to 1 VDC.

-> OUTP:LOAD INF

//Changes the output termination

 

to infinity.

-> OUTP ON

//Turns on output.

42

U2761A User’s Guide

Page 60
Image 60
Agilent Technologies U2761A manual Set Output Using SCPI Commands, Remote Interface Operation, Features and Functions