Agilent Technologies U2761A manual Start Frequency and Stop Frequency, Remote Interface Operation

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 99
Image 99
Figure 2-28Panel view of sweep

Features and Functions

2

Figure 2-28Panel view of sweep

Remote Interface Operation

SWEep:STATe {OFFON}

Start Frequency and Stop Frequency

The start frequency and stop frequency set the upper and lower frequency bounds for the sweep. The U2761A begins at the start frequency, sweeps to the stop frequency, and then resets back to the start frequency.

The range of the start and stop frequencies is 1 μHz to 20 MHz (limited to 200 kHz for Ramp and Arbitrary waveforms). The sweep is phase-continuous over the full frequency range. The default start frequency is 100 Hz. The default stop frequency is 1 kHz.

To sweep up in frequency, set the start frequency < stop frequency. To sweep down in frequency, set the start frequency > stop frequency.

Soft Front Panel Operation

On the Start Frequency and Stop Frequency panels as shown in Figure 2- 28, input the desired start frequency and stop frequency values and select the respective units from the drop down lists.

U2761A User’s Guide

81

Page 99
Image 99
Agilent Technologies U2761A Start Frequency and Stop Frequency, Remote Interface Operation, Soft Front Panel Operation