2 Features and Functions

Soft Front Panel Operation

On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list.

Remote Interface Operation

FREQuency {<frequency>}

You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Modulating Waveform

The modulating waveform consists of Sine, Square, Ramp,

Negative Ramp (Nramp), Triangle, or Arbitrary waveform.

The default modulating waveform is Sine wave.

Square has 50% duty cycle

Ramp has 100% symmetry

Triangle has 50% symmetry

Negative Ramp has 0% symmetry

Soft Front Panel Operation

Select the desired modulating waveform from the drop down

list as shown in Figure 2- 15.

Remote Interface Operation

AM:INTernal:FUNCtion

{SINusoidSQUareRAMPNRAMpTRIangleUSER}

52

U2761A User’s Guide

Page 70
Image 70
Agilent Technologies U2761A manual Modulating Waveform, Default modulating waveform is Sine wave