Agilent Technologies manual U2761A User’s Guide, Additional Information, 1Performance Criteria

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 9
Image 9
Additional Information:

Product Regulations

EMC

IEC 61326-1:2002 / EN 61326-1:1997+A1:1998+A2:2001+A3:2003

Performance Criteria

 

CISPR 11:1990 / EN 55011:1990 – Group 1 Class A

 

IEC 61000-4-2:1995 / EN 61000-4-2:1995 (ESD 4kV CD, 8kV AD)

A

IEC 61000-4-3:1995 / EN 61000-4-3:1996 (3V/m, 80% AM)

A

IEC 61000-4-4:1995 / EN 61000-4-4:1995 (EFT 0.5kV line-line, 1kV line-earth)

B

IEC 61000-4-5:1995 / EN 61000-4-5:1995 (Surge 0.5kV line-line, 1kV line-earth)

A

IEC 61000-4-6:1996 / EN 61000-4-6:1996 (3V, 0.15~80 MHz, 80% AM, power line)

A

IEC 61000-4-11:1994 / EN 61000-4-11:1994 (Dips 1 cycle, 100%)

A

Canada: ICES-001:2004

 

Australia/New Zealand: AS/NZS CISPR11:2004

 

Safety IEC 61010-1:2001 / EN 61010-1:2001

 

Canada: CAN/CSA-C22.2 No. 61010-1-04

 

USA: ANSI/UL 61010-1:2004

 

Additional Information:

The product herewith complies with the essential requirements of the Low Voltage Directive 2006/95/EC and the EMC Directive 2004/108/EC and carries the CE Marking accordingly (European Union).

1Performance Criteria:

A Pass - Normal operation, no effect.

B Pass - Temporary degradation, self recoverable.

C Pass - Temporary degradation, operator intervention required.

D Fail - Not recoverable, component damage.

N/A – Not applicable

Notes:

Regulatory Information for Canada

ICES/NMB-001:2004

This ISM device complies with Canadian ICES-001.

Cet appareil ISM est confomre à la norme NMB-001 du Canada.

Regulatory Information for Australia/New Zealand

This ISM device complies with Australian/New Zealand AS/NZS CISPR11:2004

U2761A User’s Guide

IX

Page 9
Image 9
Agilent Technologies manual U2761A User’s Guide, Additional Information, 1Performance Criteria