Agilent Technologies manual Output Frequency, Remote Interface Operation, U2761A User’s Guide

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 47
Image 47
Remote Interface Operation

Features and Functions

2

To select a function, click any of the functions on the top panel as shown below. When a function is selected, the button will be illuminated.

Figure 2-2Top panel view of the U2761A

Click Arb Output Frequency to output Arbitrary waveform. Select other Arbitrary waveform choices from the drop down list Manual background.

Remote Interface Operation

FUNCtion {SINusoidSQUareRAMPPULSeDCUSER}

You can also use the APPLy command to select the function, frequency, amplitude, and offset.

Output Frequency

As shown below, the output frequency range depends on the function currently selected. The default frequency is 1 kHz for all functions.

Table 2-2Output frequency range

Function

Minimum frequency

Maximum frequency

Sine

1 μHz

20 MHz

 

 

 

Square

1 μHz

20 MHz

 

 

 

Ramp, Triangle

1 μHz

200 kHz

 

 

 

Pulse

500 μHz

5 MHz

 

 

 

DC

Not applicable

Not applicable

 

 

 

Arbitrary

1 μHz

200 kHz

 

 

 

U2761A User’s Guide

29