Agilent Technologies FSK “Hop” Frequency, Soft Front Panel Operation, U2761A User’s Guide

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 87
Image 87
FSK “Hop” Frequency

Features and Functions

2

FSK “Hop” Frequency

The maximum alternate (or “hop”) frequency depends on the function selected as shown below. The default “hop”

frequency is 100 Hz for all functions.

Table 2-7“Hop” frequency

Function

Minimum frequency

Maximum frequency

 

 

 

Sine

1 μHz

20 MHz

 

 

 

Square

1 μHz

20 MHz

 

 

 

Ramp, Triangle

1 μHz

200 kHz

 

 

 

Arbitrary

1 μHz

200 kHz

 

 

 

Only Square wave with a 50% duty cycle is available for the Internal FSK modulating waveform.

Soft Front Panel Operation

On the Hop Frequency panel as shown in Figure 2- 22, input the desired hop frequency value and select the unit from the drop down list.

Remote Interface Operation

FSKey:FREQuency {<frequency>}

U2761A User’s Guide

69

Page 87
Image 87
Agilent Technologies U2761A manual FSK “Hop” Frequency, Soft Front Panel Operation, Remote Interface Operation