2 Features and Functions

Modulating Waveform

The modulating waveform consists of Sine, Square, Ramp,

Negative Ramp (Nramp), Triangle, or Arbitrary waveform.

The default waveform is Sine wave.

Square has 50% duty cycle

Ramp has 100% symmetry

Triangle has 50% symmetry

Negative Ramp has 0% symmetry

Soft Front Panel Operation

Select the desired modulating waveform from the drop down

list as shown in Figure 2- 20.

Remote Interface Operation

PM:INTernal:FUNCtion

{SINusoidSQUareRAMPNRAMpTRIangleUSER}

Modulating Waveform Frequency

The range of the modulating waveform frequency is 2 mHz to 20 kHz. The default modulating waveform frequency is 10 Hz.

Soft Front Panel Operation

On the Mod Frequency panel as shown in Figure 2- 20, input the desired modulating frequency value and select the unit from the drop down list.

Remote Interface Operation

PM:INTernal:FREQuency {<frequency>}