Agilent Technologies U2761A manual Modulating Waveform Frequency, Modulation Depth, N O Te

Models: U2761A

1 123
Download 123 pages 57.47 Kb
Page 71
Image 71

Features and Functions

2

Modulating Waveform Frequency

The range of the modulating waveform frequency is 2 mHz to 20 kHz. The default modulating waveform frequency is 100 Hz.

Soft Front Panel Operation

On the Mod Frequency panel as shown in Figure 2- 15, input the desired modulating frequency value and select the unit from the drop down list.

Remote Interface Operation

AM:INTernal:FREQuency {<frequency>}

Modulation Depth

 

The modulation depth is expressed in percentage and

 

represents the extent of the amplitude variation. At 0%

 

depth, the output amplitude is half of the selected value. At

 

100% depth, the output amplitude equals the selected value.

 

The range of the modulation depth is 0% to 100%. The

 

default modulation depth is 100%.

 

At 100% depth, the maximum output of the U2761A will not exceed ±2.5 V

N O TE

peak (into a 50 Ω load).

 

 

 

 

Soft Front Panel Operation

 

On the Depth panel as shown in Figure 2- 15, select the

 

desired modulating depth value from the drop down list.

 

Remote Interface Operation

 

AM:DEPTh {<depth in percent>}