Anritsu ME7840A 6-10, Receiver Display Linearity, Performance Verification, ME 7840A OMM, Step

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 106
Image 106
6-10

RECEIVER DISPLAY LINEARITY

PERFORMANCE VERIFICATION

Step 25.

Press Power key and select SOURCE 1 SETUP.

Step 26.

Set PORT 1 ATTN to 10 dB. Record the Power Meter Readout

 

and the Marker 1 Readout to the table on next page. Repeat

 

this step for other PORT 1 ATTN setting listed in the table.

Step 27.

Verify if the difference of the two measured values are within

 

specifications (Table 6-2).

Table 6-2.Receiver Magnitude Display Linearity Specifications

PORT 1 ATTN

Power Meter

MS462XC Marker 1 Reading

Difference Between Two

Specification

Setting (dB)

Reading

Readings

 

 

 

 

 

 

 

10

 

 

 

￿0.1 dB

 

 

 

 

 

20

 

 

 

￿0.1 dB

 

 

 

 

 

30

 

 

 

￿0.1 dB

 

 

 

 

 

40

 

 

 

￿0.2 dB

 

 

 

 

 

50

 

 

 

￿0.2 dB

 

 

 

 

 

60

 

 

 

￿0.2 dB

 

 

 

 

 

6-10

ME 7840A OMM

Page 106
Image 106
Anritsu ME7840A manual 6-10, Receiver Display Linearity, Performance Verification, ME 7840A OMM, Step