Anritsu manual Performance Verification, Return Loss Confidence Test, ME7840A OMM, Step, Setup

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 99
Image 99
PERFORMANCE VERIFICATION

PERFORMANCE VERIFICATION

RETURN LOSS CONFIDENCE TEST

NOTE

The worst case test result will be displayed on the screen when the instru- ment fails this test only. The failed test result is also recorded in theService Log.

Step 4.

Connect power sensor to MS462XC Port 1 and select START

 

VERIFICATION softkey.

Step 5.

Verify that Source 1 passes this test.

Step 6.

Press the SELECT SOURCE soft key to select 2

Step 7.

Connect power sensor to MS462XC Port 3 and select START

 

VERIFICATION soft key.

CAUTION

The TROUBLESHOOTING function must be properly exited to restore normal

 

Step 8.

Verify that Source 2 passes this test.

 

Step 9.

Select the menu soft key as follows to exit the TROUBLE-

 

 

SHOOTING mode:

 

 

RETURN

 

 

RETURN

 

 

FINISHED, RECOVER FROM TROUBLESHOOTINGT

 

Step 10.

Reconnect cables between MS462XC Port 1 and Port 1 of

 

 

MS4782 Test Set and between Port 3 of MS462XC and Port 3

 

 

of MS4782 Test Set.

6-5RETURN LOSS

This test verifies the accuracy of return loss measurements.

CONFIDENCE

 

 

TEST

 

 

Setup

Turn on the MS462XC VNMS and MS4782X Test Set and allow them to

 

warm up for 30 minutes.

 

Step 1.

Press Default key then 0 key.

 

Step 2.

Press Seq key and .select:

 

 

TTL I/O.

 

 

PARALLEL OUT SETUP

 

 

SET PARALLEL OUT PORT (0-255)

 

Step 3.

Change the value from 0 to 8.

 

Step 4.

Select:

ME7840A OMM

6-3

Page 99
Image 99
Anritsu manual Performance Verification, Return Loss Confidence Test, ME7840A OMM, Step, Setup