OPERATIONS, MEASUREMENT | POWER SWEEP, TWO TONE |
Return: Returns to the main PATS screen.
Change Params: Returns to User Input
screen.
Y1, Y2: Provides for rescaling the left or right axis.
Rescale: Rescales the measurement results.
Max/Min: Type in maximum and minimum display value. Use Rescale after changing val- ues.
Markers Screen Options (left):
Markers slider: Moves marker between data
points.
Gain dB: Gain (or S21) is measured at each CW tone as the power is swept The measure- ment is made in IMD mode. The Gain is actu- ally measured by calculating the difference be- tween Power In (in dBm) and Power Out (in dBm). Calibration applied is Linear Power Cal and Receiver Cal. (Could also be called Gain and compression measurement). The displayed Power In level is the power of each input tone.
Gain dB: See above.
IM3U: Intermodulation product 3 upper side- band
IM3L:Intermodulation product 3 lower side- band
IM5U:Intermodulation product 5 upper side- band
IM5L:Intermodulation product 5 lower side- band
IM7U:Intermodulation product 7 upper side- band
IM7L:Intermodulation product 7 lower side- band
PAE:
Drain: Drain current value.
ME7840A OMM |
|