POWER LEVEL

POWER LEVEL

OPERATIONS, CALIBRATION

sure that the equipment is properly connected to the bus. If so, the following screen will be displayed:

Step 3. Click on “Run Calibration Procedure” to display a list of cali- bration specification files (below). (Refer to paragraph 4-3for a description of these files.) Running the files is described in succeeding numbered paragraphs.

4-6POWER LEVEL

The A_Power Level.txt specification file provides prompts for a power

 

level calibration. Proceed as follows:

 

Step 1.

Zero-set the power sensor for accurate measurement in fol-

 

 

lowing steps.

 

Step 2.

Select “Tools” then “Run Calibration Procedure.”

4-4

ME 7840A OMM

Page 46