INTRODUCTION

This chapter describes the five measurements available with the

 

ME7840A software: S-Parameters, Power Sweep, IMD (intermodulation

 

distortion), Harmonics, and Hot S22.

OPERATION,

Refer to Chapter 3 for general operation and setup of the ME7840A.

GENERAL

 

MEASUREMENT

Measurements always include a degree of uncertainty due to imperfec-

CALIBRATION

tions in the measurement system. The measured value is always a combi-

 

nation of the actual value plus the systematic measurement errors. Cali-

 

bration, as it applies to network analysis, characterizes the systematic

 

measurement errors and subtracts them from the measured value to ob-

 

tain the actual value. Each of the ME7840A measurements requires a cal-

 

ibration to account for measurement uncertainties. Refer to Chapter 4 for

 

measurement calibration procedures.

5-4GENERAL
Calibration(s)

Completed

Measurements can be approached from two directions: calibration(s) com- pleted, calibration(s) not completed.

Proceed as follows:

Step 1. Click on “Tools,” and open “Check Communications to VNA.”

Step 2. Verify the communications are correct.

Step 3. Click on “Tools,” and open “Calibration File Options.”

Step 4. Click on “Find and Recall” and recall a *.cal file for each mod- ule. Filenames are listed below and shown in graphic below last listing.

￿S Parameters: “S_Parameters_1.cal”

￿Power Sweep (2T): “PowSweep_1.cal”

￿Power Sweep (1T): “PSweep1T_1.cal”

￿C3, IMD: “Imd_1.cal”

￿C4, Harmonics: “Harmonics_1.cal”

ME7840A OMM

5-1