GENERAL

GENERAL

OPERATIONS, MEASUREMENT

￿C5, Hot S22 ￿ “Hots22_1.cal”

Step 5. Select appropriate file “Open” (below).

Calibration(s) Not Sometimes users may wish to run a measurement using default settings,

Completedthat is, without spending the time required to calibrate. The procedure below gives instructions.

Step 1. Click on “Tools,” and open “Check Communications to VNA.”

Step 2. Verify the communications are correct.

Step 3. Click on “Tools,” and open “Calibration File Options.”

Step 4. Clear the text for all modules then select “Return.”

Step 5. Press any measurement button on main menu and measure- ment will be run using “PATS software” default values: Cali- bration Off on MS462XC and MS462XC Default parameters in place. The default values are shown below.