Anritsu ME7840A manual Technical Details, Cal Specification Files, ME 7840A OMM

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 118
Image 118
Manual background

TECHNICAL DETAILS

CAL SPECIFICATION FILES

Table A-3.Technical Details of Calibration Specification Files (7 of 9).

3.powsweep (Power Sweep ￿ IMD mode or 2-tone mode):

POWSWEEP_START Header

Details

POWSWEEP_END

Header Section Keywords:

ncal

Number of calibration sequences of this type to run.

npts

NOTE: npts is IGNORED. Always uses 3, which refers to number of frequency values

 

allowed (must be 3 in current software).

if

IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000

 

NOTE: Illegal value causes software to default to 1000.

avgs

Averages: (Optional). Software will set averaging to value (and turn on) during cal.

pow

Start power level for sweep: Set on for Source 1 and Source 2.

switch

Setting for test set. Correponds to 0￿15 settings.

Detail Section Keywords:

CAL, 1

First calibration

name

Name of *.cal file to save when calibration sequence completes.

f1

Start frequency in MHz for global calibration

 

NOTE: Set a wide frequency range. Miminum value is 10; ensure f1 is less

 

than 2.

f2

Stop frequency in MHz.

Example

Refer to Figure A-2(page A-12).

A-8

ME 7840A OMM

Page 118
Image 118
Anritsu ME7840A manual Technical Details, Cal Specification Files, ME 7840A OMM