CAL FILE SETUP

OPERATIONS, CALIBRATION

CAL FILE SETUP

4-9CAL FILE SETUP

The “D_Cal File Setup.txt” file provides prompts for setting up One-tone

 

power sweep, IMD, Two-Tone Power Sweep, and Harmonic specification

 

file. The prompts cause the Scorpion to be placed into a state where it is

 

ready to make calibrated measurements on the user’s AUT. At the com-

 

pletion of the program, the state information is saved to a *.cal file. To

 

run this program, proceed as follows.

 

Step 1.

Select “Tools” then “Run Calibration Procedure.”

 

Step 2.

Double-click on “B_Cal Files Setup.txt.” to begin with the file

 

 

setup for a Transmission Frequency Response (S21 cal) for

 

 

One-Tone Power Sweep. Connect a throughline when di-

rected in a following step.

ME7840A OMM

4-15