Anritsu manual S-Parametertest K Factor, Operations, Measurement, ME7840A OMM, Test K Factor

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 68
Image 68
S-PARAMETER TEST: K FACTOR

OPERATIONS, MEASUREMENT

S-PARAMETER TEST: K FACTOR

5-6S-PARAMETER

This test describes the K Factors measurement. This measurement is

TEST: K FACTOR

described in Chapter 1, paragraph 1-10.

 

 

 

 

Procedure.

 

 

 

 

 

Prepare the ME7840A as shown in Chapter 3.

 

 

Enter the appropriate test values in the displayed

 

 

User Input screen. The input fields are described

 

 

below: The inputs required for this test are de-

 

 

scribed below.

 

 

 

 

 

￿ S-Parameter:Drop down options. K Factor for

 

 

 

this measurement.

 

 

 

 

￿ Data Points: Drop down options ￿ number of data

 

 

 

points in frequency sweep.

 

 

 

￿ Frequency 1 (MHz): Lower frequency or start fre-

 

 

 

quency.

 

 

 

 

 

￿ Frequency 2 (MHz): Upper frequency or stop fre-

 

 

 

quency.

 

 

 

 

 

￿ Input Power (dBm):Input power level, in dBm.

 

 

￿ IF Bandwidth (Hz):Intermediate frequency band-

 

 

 

width for test, in Hertz.

 

 

 

 

￿ Test Set Setting: Byte value for setting test set

 

 

 

attenuation and source switch (below).

 

 

 

 

 

 

 

 

 

 

 

 

Internal Source

 

Value

External Source

Value

 

 

 

 

 

 

 

 

 

 

 

0 dB

 

8

0 dB

0

 

 

 

 

 

 

 

 

 

 

 

10 dB

 

9

10 dB

1

 

 

 

 

 

 

 

 

 

 

 

20 dB

 

10

20 dB

2

 

 

 

 

 

 

 

 

 

 

 

 

 

30 dB

 

11

30 dB

3

 

 

 

 

 

 

 

 

 

 

 

40 dB

 

12

40 dB

4

 

 

 

 

 

 

 

 

 

 

 

50 dB

 

13

50 dB

5

 

 

 

 

 

 

 

 

 

 

 

60 dB

 

14

60 dB

6

 

 

 

 

 

 

 

 

 

 

 

70 dB

 

15

70 dB

7

 

 

 

 

 

 

 

 

ME7840A OMM

5-7

Page 68
Image 68
Anritsu manual S-Parametertest K Factor, Operations, Measurement, ME7840A OMM, Test K Factor, Procedure