TECHNICAL DETAILS

CAL SPECIFICATION FILES

Table A-1.Technical Details of Calibration Specification Files (5 of 9).

2.sp and Hot S22 (S Parameters & Hot S22)

SP_START Header Details

SP_END

Header Section Keywords:

ncal

Number of calibration sequences of this type to run.

npts

Number of frequency points to use during the calibration sequence. Increasing

 

this number will increase calibration time.

 

Legal values are (3, 15, 51, 101, 201, 401, 801, 1601 only)

 

NOTE: Illegal value causes software to default to 51.

con

Connector sypes for Port 1 and Port 2: Options:

 

MN, FN - Male/Female N type

 

MK, FK - Male/Female K

 

MS, FS - Male/Female SMA

if

IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000

 

NOTE: Illegal value causes software to default to 1000.

avgs

Averages: (Optional). Software will set averaging to value (and turn on) during cal.

pow

Source power setting: during cal, Source 1.

switch

Setting for test set. Correponds to 0￿15 settings.

Detail Section Keywords:

CAL, 1

First calibration

cal type

Type of calibration to perform:

 

12 - 12 term

 

RR - Reflection Reverse

 

RF - Reflection Forward

 

8T - 1 Path 2 Port Forward

 

8R - 1 Path 2 Port Reverse

 

FT - Transmission Frequency Response

<name>

Name for Hot S22

<hot>

For Hot S22 only. Value = offset in kHz of 2nd Source

 

 

A-6