S-PARAMETERS CAL

OPERATIONS, CALIBRATION

S-PARAMETERS CAL

4-8S-PARAMETERS

The “C_S-Parameters Cal.txt” file to provides prompts for setting up

CAL

S-parameters specification file. The prompts cause the Scorpion to be

 

placed into a state where it is ready to make calibrated measurements on

 

the user’s AUT. At the completion of the program, the state information is

 

saved to a *.cal file. To run this program, proceed as follows.

 

Step 1.

Select “Tools” then “Run Calibration Procedure.”

 

Step 2.

Double-click on “D_S Parameters.txt” to begin the file

 

 

setup.Follow the prompt and connect open, short, and load to

 

 

Ports 1 and 2 as directed.

 

 

 

ME7840A OMM