HARMONICS

OPERATIONS, MEASUREMENT

HARMONICS

5-10HARMONICS

The harmonics test measures and displays the harmonics in the device

 

under test.

 

 

 

 

Procedure

 

 

 

 

 

Prepare the ME7840A as shown in Chapter 3.

 

 

Enter the appropriate test values in the displayed

 

 

User Input screen. The input fields are described

 

 

below:

 

 

 

 

 

￿ Measurement Type: Already set to “HA” for Har-

 

 

 

monics – input disabled.

 

 

 

 

￿ Start Frequency (MHz): Lower frequency to be

 

 

 

displayed. Typically set to a value less than or

 

 

 

equal to the main tone.

 

 

 

 

￿ Stop Frequency (MHz): Upper Frequency to be

 

 

 

displayed. Set to a value less than the system’s

 

 

 

maximum frequency (for example, for MS4623C

 

 

 

set to a max of 5000 MHz).

 

 

 

￿ Main Tone (MHz): CW Frequency setting for

 

 

 

source 1.

 

 

 

 

 

￿ Input Power: Input RF power to AUT.

 

 

 

￿ Test Set Setting: Byte value for setting test set

 

 

 

attenuation and source switch:

 

 

 

 

 

 

 

 

 

 

 

Internal Source

Value

External Source

Value

 

 

 

 

 

 

 

 

 

 

0 dB

8

0 dB

0

 

 

 

 

 

 

 

 

 

 

10 dB

9

10 dB

1

 

 

 

 

 

 

 

 

 

 

 

 

20 dB

10

20 dB

2

 

 

 

 

 

 

 

 

 

 

30 dB

11

30 dB

3

 

 

 

 

 

 

 

 

 

 

40 dB

12

40 dB

4

 

 

 

 

 

 

 

 

 

 

50 dB

13

50 dB

5

 

 

 

 

 

 

 

 

 

 

60 dB

14

60 dB

6

 

 

 

 

 

 

 

 

 

 

70 dB

15

70 dB

7

 

 

 

 

 

 

 

ME7840A OMM

5-29

Page 90
Image 90