Anritsu manual Power Leve.Txt, A-11, Cal Specification Files, ME7840A OMM, A-5 POWER LEVE.TXT

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 121
Image 121
POWER LEVE.TXT

CAL SPECIFICATION FILES

POWER LEVE.TXT

A-5POWER LEVE.TXT

The Power Level.txt file is shown in Figure A-1

 

 

ndut,1,

//Number of DUTs

type,pow,

//Power Level Calibration (requires Power Meter)

 

//Perform Global Linear, Flat Power and Receiver

 

Calibrations

POW_START ncal,1,

//Number

of Power Level calibrations

npts,401,

//Number

of points (default for all)

if,3000,

//IF default

avgs,0,

//averages default

pow,0,

//power level default

switch,8,

//switch

setting

CAL,1, name,Levels,

//File name root (append _1 for DUT1, _2 for DUT2, etc

f1,100,

//startf

DUT1 DUT2 DUT3

f2,3000,

//stopf

DUT1 DUT2 DUT3 POW_END

END

Figure A-1.Power Level Calibration Specification File

A-6CAL FILES

The Cal Files Setup.txt file is shown in Figure A-2.

SETUP.TXT

 

ME7840A OMM

A-11

Page 121
Image 121
Anritsu manual Power Leve.Txt, A-11, Cal Specification Files, ME7840A OMM, A-5 POWER LEVE.TXT, A-6 CAL FILES, Setup.Txt