HOT S22.TXT

CAL SPECIFICATION FILES

HOT S22.TXT

A-7HOT S22.TXT

The HotS22.txt file is shown in Figure A-3.

 

 

ndut,1,

//Number

of

DUTs

type,sp,

//S Parameter calibration

SP_START

 

 

 

ncal,1,

//Number

of

S-Param calibrations

npts,201,

//Number

of

points (default for all)

con,FN,FN,

//Connector

Type Options MN,FN,MK,FK,MS,FS

if,3000,

//IF default

 

avgs,0,

//averages default

pow,-5,

//power level default

switch,8,

//switch

setting

CAL,1, caltype,RR,

//12 = 12 Term, RR = Relection only Reverse, RF =

 

Relection

only Forward

name,Hots22,

//File name

root (append _1 for DUT1, _2 for DUT2, etc

hot,923,

//hot 0

means turn on 2nd source during cal (Hot S22

 

only),

 

 

 

//use value

as offset (KHz)

pow2,-5,

//2nd source power in Hot S22

f1,800,

//startf

DUT1 DUT2 DUT3

f2,1000,

//stopf

DUT1 DUT2 DUT3

SP_END

END

Figure A-3.Hot S22 Calibration Specification Program File