Anritsu ME7840A System Dynamic Range, Performance Verification, ME 7840A OMM, Test Procedure

Models: ME7840A

1 126
Download 126 pages 42.52 Kb
Page 102
Image 102
SYSTEM DYNAMIC RANGE

SYSTEM DYNAMIC RANGE

PERFORMANCE VERIFICATION

Table 6-1.Return Loss Specifications

Offset Termination

Measured S11 Value

Measured S22 Value*

Specification

 

 

 

 

SC5237

 

 

6 ￿0.4 dB

 

 

 

 

SC5270

 

 

20 ￿1.0 dB

 

 

 

 

* Apply to Test Set with Hot S22 Capability only.

6-6SYSTEM DYNAMIC

This test verifies the PATS system dynamic range.

RANGE

 

Test Procedure:

Step 1. Press Default key then 0 key of MS462XC to reset the system.

Step 2. Press Seq key.

Step 3. Select:

TTL I/O.

PARALLEL OUT SETUP

SET PARALLEL OUT PORT (0-255)

Step 4. Change the value from 0 to 8.

Step 5. Select:

RETURN

RETURN

Step 6. Press Ch 3 key.

Step 7. Press Display key and select:

DISPLAY MODE

SINGLE CHANNEL

RETURN

GRAPH TYPE

LOG MAGNITUDE

RETURN

Step 8. Press Cal key.

Step 9. Follow the prompts and select::

6-6

ME 7840A OMM

Page 102
Image 102