Section 8 Device Message Details

:STATus : OPERation : INSTrument : PTRansition?

Command type

Sequential command

Response

<numeric> = <NR1 NUMERIC RESPONSE DATA>

 

0 to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Makes a query of the details of the transition filter (a change in the positive direction) of the

 

INSTrurnent status register.

Example of use

When the bit of measurement completion is set at PTRansition

>: STATus : OPERation : INSTrument : PTRansition?

<2

:STATUS:OPERation : INSTrument : NTRansition <numeric>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

O to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Sets the transition filter (a change in the negative direction) of the INsTmment status register

 

Set at I .

Example of use

Sets the bit of measurement completion at NTRansition.

>: STATUS: OPERation : INSTrurnent : NTRansition 2

:STATus : OPERation : INSTrument : NTRansition?

Command type

Sequential command

Response

< n u m e r i o = INRI NUMERIC RESPONSE DATA>

 

O to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Makes a query of the details of the transition filter (a change in the negative direction) of the

 

INSTrurnent status register.

Example of use

When the bit of measurement completion is set at NTRansition.

 

> : STATus : OPERation : INSTrument : NTRansition'?

 

< 2

Page 110
Image 110
Anritsu MP1777A operation manual STATUSOPERation INSTrument NTRansition numeric, STATus OPERation INSTrument NTRansition?