Section 8 Device Message Details
:STATus : OPERation : INSTrument : PTRansition?
Command type | Sequential command |
Response | <numeric> = <NR1 NUMERIC RESPONSE DATA> |
| 0 to 32,767 ........ Sum (decimal digit) of transition filter bits |
Function | Makes a query of the details of the transition filter (a change in the positive direction) of the |
| INSTrurnent status register. |
Example of use | When the bit of measurement completion is set at PTRansition |
>: STATus : OPERation : INSTrument : PTRansition?
<2
:STATUS:OPERation : INSTrument : NTRansition <numeric>
Command type | Sequential command |
Parameter | <numeric> = <DECIMAL NUMERIC PROGRAM DATA> |
| O to 32,767 ........ Sum (decimal digit) of transition filter bits |
Function | Sets the transition filter (a change in the negative direction) of the INsTmment status register |
| Set at I . |
Example of use | Sets the bit of measurement completion at NTRansition. |
>: STATUS: OPERation : INSTrurnent : NTRansition 2
:STATus : OPERation : INSTrument : NTRansition?
Command type | Sequential command |
Response | < n u m e r i o = INRI NUMERIC RESPONSE DATA> |
| O to 32,767 ........ Sum (decimal digit) of transition filter bits |
Function | Makes a query of the details of the transition filter (a change in the negative direction) of the |
| INSTrurnent status register. |
Example of use | When the bit of measurement completion is set at NTRansition. |
| > : STATus : OPERation : INSTrument : NTRansition'? |
| < 2 |