Section 8 Device Message Details

:SENSe : TELecom : MEASure : PERiod <numeric>, <suffix>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

1 to 99

 

 

<suffix> = <CHARACTER PROGRAM DATA>

 

EI

hour

 

M

minute

 

S

second

Function

Performs the measurement time setting.

Restriction

The setting is disabled when the measurement mode is set to manual measurement.

Example of use

Sets the measurement time at one hour.

>: SENSe : TELecom : MEASure : PERiod 1 , H

:SENSe : TELecom : MEASure : PERiod?

Command type

Sequential command

Response

<numeric>, <suffix>

 

<numeric> = <NRl NUMERIC RESPONSE DATA>

 

<suffix> = <CHARACTER RESPONSE DATA>

Function

Makes a query of the measurement time.

Example of use

> : SENSe : TELecom : MEASure : PERiod?

 

< 1, H