Section 8 Device Message Details
8.1Response Format
This section explains the format of the rehponse to the query command
The format is shown in the table below.
Table 8-1 Response Format (By Response Type)
Response Type
<numeric>, <year>
<NRl NUMERIC
RESPONSE DATA>
<brate>, <type>...etc
<CHARACTER
RESPONSE DATA>
I
<string>, <display>...etc
<STRING RESPONSE
DATA>
/
Format
Makes the number of digits of the response variable and sets the maximum number of digits within the range of numerical values as the maximum number of digits. No space is inserted between the sign and numerical value.
>: SYSTem : DATE?
i1993.7, 14
A short form of a character is returned.
>: INSTrument : COUPle?
<ALL
A string enclosed by the quotation marks (") is returned
When short and long forms are both found in the string of the corresponding command, the short form is returned.
>: DISPiay : DSELect : NAME?
<"SET" (short form of "SETUP")
See Table