8.3 
:STATus : OPERation : PTRansition <numeric>
| Command type | Sequential command | 
| Parameter | < n u m e r i o = <DECIMAL NUMERIC PROGRAM DATA> | 
| 
 | 0 to 32,767 ........ Sum (decimal digit) of transition filter bits | 
| Function | Sets the transition filter (a change in the positive direction) of the OPERation status register. | 
| 
 | Set at I . | 
| Example of use | Sets the bit being measured at PTRansition. | 
>: STATus : OPERation : PTRansition 16
:STATus : OPERation : PTRansition?
| Command type | Sequential command | 
| Response | <numeric> = <NR 1 NUMERIC RESPONSE DATA> | 
| 
 | 0 to 32,767 ........ Sum (decimal digit) of transition filter bits | 
| Function | Makes a query of the details of the transition filter (a change in the positive direction) of the | 
| 
 | OPERation status register. | 
| Example of use | When the bit being measured is set at PTRansition | 
| 
 | > : STATUS: OPERation : PTRansition? | 
| 
 | < 16 | 
- - - ---
:STATUS: OPERation : NTRansition <numeric>
| Command type | Sequential command | 
| Parameter | <numeric> = <DECIMAL NUMERIC PROGRAM DATA> | 
| 
 | 0 to 32,767 ........ Sum (decimal digit) of transition filter bits | 
| Function | Sets the transition filter (a change in the negative direction) of the OPERstion status register. | 
| 
 | Set at 1. | 
| Example of use | Sets the INSTrument status register summary at NTRansition. | 
| 
 | > : STATus : OPERation : NTRansition 8 192 | 