Anritsu MP1777A STATus OPERation PTRansition?, Status OPERation NTRansition numeric

Models: MP1777A

1 130
Download 130 pages 12.22 Kb
Page 107
Image 107

8.3 Device-Specific Commands

:STATus : OPERation : PTRansition <numeric>

Command type

Sequential command

Parameter

< n u m e r i o = <DECIMAL NUMERIC PROGRAM DATA>

 

0 to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Sets the transition filter (a change in the positive direction) of the OPERation status register.

 

Set at I .

Example of use

Sets the bit being measured at PTRansition.

>: STATus : OPERation : PTRansition 16

:STATus : OPERation : PTRansition?

Command type

Sequential command

Response

<numeric> = <NR 1 NUMERIC RESPONSE DATA>

 

0 to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Makes a query of the details of the transition filter (a change in the positive direction) of the

 

OPERation status register.

Example of use

When the bit being measured is set at PTRansition

 

> : STATUS: OPERation : PTRansition?

 

< 16

- - - ---

:STATUS: OPERation : NTRansition <numeric>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

0 to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Sets the transition filter (a change in the negative direction) of the OPERstion status register.

 

Set at 1.

Example of use

Sets the INSTrument status register summary at NTRansition.

 

> : STATus : OPERation : NTRansition 8 192

Page 107
Image 107
Anritsu MP1777A operation manual STATus OPERation PTRansition?, Status OPERation NTRansition numeric