Section 8 Device Message Details
:SENSe : TELecom : RANGe <range>
Command type Parameter
Function Example of use
Sequential command
<range> = <CHARACTER PROGRAM DATA>
| U14 | ..............4 UIPP | 
| UI 1 | ..............1 UIPP | 
Sets the measurement range of receiving jitter.
Sets the measurement range of receiving jitter at 1 UIPP,
>: SENSe : TELecom : RANGe UIl
:SENSe : TELecom : RANGe?
| Command type | Sequential command | 
| Response | <range> = <CHARACTER RESPONSE DATA> | 
| Function | Makes a query of receiving jitter measurement. | 
| Example of use | > : SENSe : TELecom : RANGe? | 
| 
 | < UII | 
: SENSe :TELecom : FlLTer <filter>
| Command type | Sequential command | 
 | |
| Parameter | <filter> = <CHARACTER PROGRAM DATA> | ||
| 
 | LP | LP | (80 M) | 
| 
 | H P l L P | HPl+LP | (10 K to 80 M) | 
| 
 | HPl SLP | HPI1+LP | (20 K to 80 M) | 
| 
 | HP2LP | HP2+LP | (4 M to 80 M) | 
| 
 | HP3LP | HP3+LP | (6 M to 80 M) | 
| 
 | HPLP | HP+LP | ( 12 K to 80 M) | 
| 
 | HPSLP | HP'+LP | (50 K to 80 M) | 
| Function | Performs filter setting. | 
 | |
| Restriction | H P 1 + L P , H P l ' + L P setting is disabled w h e n the Bit Rate is set a t 2.5 G o r 5 G. | ||
| 
 | HP3+LP can be set only for Option 07. | ||
| Example of use | Sets the filter at HP'+LP, HPll+LP. | ||
| 
 | > : SENSe : TELecom : FILTer HPSLP | ||