Section 8 Device Message Details

:SYSTern : MEMory : CLEar <numeric>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

1 to 10

Memory number 1 to Memory number 10

Function Restriction Example of use

Deletes set data in memory.

Specifying a memory number where no data is written will result in a run time error.

Deletes data of memory number 3.

>: SYSTern : MEMory : CLEar 3

:SYSTern : MEMory : LABel <numeric>, <title>

Command type Parameter

Function Example of use

Sequential command

<numeric> = <DECIMAL NUMERIC PROGRAM DATA> 1 to 1 0 ......... Memory number 1 to Memory number 10

<title> = <STRING PROGRAR.1DATA>

A maximum of 15 characters (excluding the quotation marks) "Name of memory"

Writes in the name of memory for set data.

Sets the name of memory number 1 at "2488 M test."

>: SYSTem : MEMory : LABel 1, "2488 M test"

:SYSTern : MEMory : LABel? <numeric>

Command type Parameter

Function Example of use

Sequential command

<numeric> = <DECIh?AL NUMERIC PROGRAM DATA> 1 to 10 ......... Memory number 1 to Memory number 10

Response <title> = <STRING RESPONSE DATA>

A maximum of 15 characters (excluding the quotation marks)

Makes a query of the name of the memory for set data.

Makes a query of the name of memory number 1.

>: SYSTem : MEMory : LABel? 1

<"2488 M test"

Page 102
Image 102