8.3 Device-Specific Commands

:STATus :OPERation : INSTrument : ENABle <numeric>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

0 to 32,767 ........ Sum (decimal digit) of event enable register bits

Function

Sets the masking value of the event enable register of the INSTrument register.

 

Masking at 0.

Example of use

Sets the masking value in a manner that when the bit of alarm change of the event register

 

switches ON, it will be reported to the OPERation register.

>: STATus : OPERation : INSTrument : ENABle 16

:STATus : OPERation : INSTrument : ENABle?

Command type

Sequential command

Response

<numeric> = <NR1 NUMERIC RESPONSE DATA>

 

0 to 32,767 ........ Sum (decimal digit) of event enable register bits

Function

Makes a query of the details of the event enable register of the INSTrument register.

Example of use

When only the bit of alarm change is enabled.

 

> : STATus : OPERation : INSTrument : ENABle?

 

< 16

--

:STATUS: OPERation :INSTrument : PTRansition <numeric>

Command type

Sequential command

Parameter

<numeric> = <DECIMAL NUMERIC PROGRAM DATA>

 

O to 32,767 ........ Sum (decimal digit) of transition filter bits

Function

Example of use

Sets the transition filter (a change in the positive direction) of the INSTrument status register. Set at 1.

Sets the bit of measurement completion at PTRansition.

>: STATus : OPERation : INSTrument : PTRansition 2