Instruction Manual
Model NGA 2000 TO2
|
|
a. General |
|
Measurement Species | Trace Oxygen |
Ranges | 0 to 100 ppm (output scalable down to |
Accuracy | ±3% of reading or ±0.02% of range (except for ranges ≤ |
| 100 ppm: ±3% of reading or ±0.05% of range) |
Sensitivity | <10 ppb Oxygen |
Noise | 1% of fullscale, peak to peak |
Linearity | ±1% of fullscale |
Response Time | Typically 90% in less than 20 seconds |
Zero Drift | ≤±1% of fullscale/24 hours at constant temperature |
Span Drift | ≤±1% of fullscale/24 hours at constant temperature |
Effect of Temperature | 0.32% of reading per °F from 70°F (0.58% of reading per |
| °C from 21°C) |
Effect of Flow | ≤2% of reading for a flow change of ±250 cc/min (0.5 |
| SCFH) |
Operating Temperature | 32°F to 113°F (0°C to 45°C) |
Power Requirements | +24 VDC ±5%, 10 W max. |
Ripple and Noise | <100 mV peak to peak |
Line and Load Regulations | <±1% |
b. Sample |
|
Sample | |
Flow Rate | 0.5 to 1.5 L/min. |
Supply Pressure | 1027 to 1082 hPa - absolute (0.2 to 1.0 psig) |
Temperature | 32°F to 113°F (0°C to 45°C) |
Particulates | filtered to <0.1 mg/L; |
| perature |
Sample Humidity |
c. Physical |
|
Materials in contact with sample | Stainless steel, Teflon, Delrin, neoprene |
Dimensions | See Figure |
Weight | 6.8 kg (15 lbs.) |
Mounting | Horizontal, external to Platform or custom installed in a |
| panel |
Case Classification | General Purpose for installation in weather protected |
| area |
Max. Separation from Platform | 1600 m (1 mile) |
1See the Platform manual for specifications regarding Platform related components.
Description and Specifications | Rosemount Analytical Inc. A Division of Emerson Process Management |