Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

1-5 SPECIFICATIONS1

 

a. General

 

Measurement Species

Trace Oxygen

Ranges

0 to 100 ppm (output scalable down to 0-2 ppm fullscale)

Accuracy

±3% of reading or ±0.02% of range (except for ranges

 

100 ppm: ±3% of reading or ±0.05% of range)

Sensitivity

<10 ppb Oxygen

Noise

1% of fullscale, peak to peak

Linearity

±1% of fullscale

Response Time

Typically 90% in less than 20 seconds

Zero Drift

≤±1% of fullscale/24 hours at constant temperature

Span Drift

≤±1% of fullscale/24 hours at constant temperature

Effect of Temperature

0.32% of reading per °F from 70°F (0.58% of reading per

 

°C from 21°C)

Effect of Flow

2% of reading for a flow change of ±250 cc/min (0.5

 

SCFH)

Operating Temperature

32°F to 113°F (0°C to 45°C)

Power Requirements

+24 VDC ±5%, 10 W max.

Ripple and Noise

<100 mV peak to peak

Line and Load Regulations

<±1%

b. Sample

 

Sample

Non-flammable (below 100% of the LEL)

Flow Rate

0.5 to 1.5 L/min.

Supply Pressure

1027 to 1082 hPa - absolute (0.2 to 1.0 psig)

Temperature

32°F to 113°F (0°C to 45°C)

Particulates

filtered to <0.1 mg/L; non-condensing at ambient tem-

 

perature

Sample Humidity

non-condensing at ambient temperatures

c. Physical

 

Materials in contact with sample

Stainless steel, Teflon, Delrin, neoprene

Dimensions

See Figure 2-2, Outline and Mounting Dimensions

Weight

6.8 kg (15 lbs.)

Mounting

Horizontal, external to Platform or custom installed in a

 

panel

Case Classification

General Purpose for installation in weather protected

 

area

Max. Separation from Platform

1600 m (1 mile)

1See the Platform manual for specifications regarding Platform related components.

1-4

Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 16
Image 16
Emerson NGA 2000 TO2 instruction manual SPECIFICATIONS1, General, Sample, Physical