Model NGA 2000 TO2

Instruction Manual

760008-A September 2001

b.Alarms

From the Analyzer Module Setup and Controls menu, choose Alarms as follows:

Main Menu

Analyzer and I/O, expert controls and setup…

TO2

37.50 ppm

 

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

Status…

Back…

TO2

37.50 ppm

-- Alarms --

 

Alarm delay:

1.0 s

Concentration alarm setup…

 

Measure

Back…

ClrAlm!

Move the cursor to the “Alarms…” line and press the key. The “Alarms” menu will dis- play as shown below.

The “Alarm delay” function on the Alarms menu establishes the time between an ex- ceedance and the triggering of the alarm. It is adjustable from 0.0 to 30.0 seconds. Move the cursor to the “Alarm delay:” line and press the key. Adjust the value using the and keys or select the digit using the and keys.

Use the F5 (ClrAlm!) key to clear any acti- vated alarms. This should be done only after the signal has returned to the allowed range to reset the alarm for new events.

Move the cursor to the “Concentration alarm setup..” line and press the key to reach the “Concentration Alarm Setup” menu as shown below.

TO2

 

37.50 ppm

 

 

 

-- Concentration Alarm Setup --

 

 

Alarm generation is:Re

 

s…On

 

 

Level for Low-Low alarm:

 

0.0 ppm

 

 

Level for Low alarm:

 

10.0 ppm

 

 

Level for High alarm:

 

50.0 ppm

 

 

Level for High-High alarm:

 

100.0 ppm

 

 

Low-Low alarm:

 

Off

 

 

Low alarm:

 

On

 

 

High alarm:

 

Off

 

 

High-High alarm:

 

Off

 

 

Measure

 

Back…

 

 

To turn alarms on or off, select the “Alarm generation is:” line and toggle it between “On” or “Off.”

Four alarms are available. Select any alarm and adjust the concentration value for trig- gering that alarm. Setting an alarm concen- tration to 0.0 will deactivate that alarm.

The status of the alarms displays below the line.

Figure 3-18. Alarms