Instruction Manual
Model NGA 2000 TO2
Date and Time
This menu is used to set the date, time and format for the analyzer.
TO2 | 37.50 ppm |
|
CAUTION.
If a security code is lost or forgotten, there is no possibility of entering the locked se- curity level.
Minutes:
Hours:
Year:
Day:
Month:
Network updating: Current time:
Measure
0
12
2000
1
2
Enabled
08:45:35 February 01, 2000
Set! Back…
Setting the code numbers:
Enter the submenu for the desired secu- rity level to set the PIN. The default val- ues are:
Basic level: 12345
Select a line with the ↑ or ↓ keys.
Press the ↵ or → keys to select the pa- rameter.
Select any digit with the ← or → key and set a new value with the ↑ or ↓ key.
Expert level: 54321
System level: 12345
Use the function keys F1 to F5 to enter the numerical security code in the desired sequence. The numbers will appear in the “Actual PIN” line as they are entered. The characters displayed on the function keys cannot be entered as code numbers.
Network updating:
Not used with this model.
Set up a new date or time:
Set the “Minutes,” “Hours,” “Year,” “Day,” or “Month” lines and make any desired adjustments. Press the F3 key to set the new time and date.
The “Current time” line will change to re- flect the new time and date set.
Security Codes
TO2 |
| 37.50 ppm | ||||
|
|
|
| |||
|
|
|
| |||
| Basic level security: |
| Disabled |
| ||
| Expert level security: |
| Disabled | |||
| System level security: |
| Disabled | |||
| Define basic level security PIN… |
|
|
| ||
| Define expert level security PIN… |
|
|
| ||
| Define system level security PIN… |
|
|
| ||
| Measure |
|
| Back… |
|
|
This menu is used to set the security codes for the three levels of security.
Example:
TO2 | 37.50 ppm |
|
Press five softkeys in any order to define the PIN
‘
The actual PIN is represented by the order in which they are
pressed, and shown numerically below.
Press the left arrow key when you are done.
Actual PIN: |
| 12345 | ||||||
ABCDE1 |
| FGHIJ2 |
| KLMNO3 |
| PQRST4 |
| UVWXYZ5 |
|
|
|
|
Press the ← key to return to the “Security setup” menu.
Enable the security code
Select the desired security level line to enable. Change the parameter to “En- abled.”
CAUTION.
If System level is enabled, it will not be possible to
Rosemount Analytical Inc. A Division of Emerson Process Management |