Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

g.Flow Measurement

Sample flow should be maintained within a range suitable for the desired accuracy. See Effect of Flow in Section 1-5a.

Main Menu

Analyzer and I/O, expert controls and setup…

TO2

37.50 ppm

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

Status…

Back…

TO2

37.50 ppm

 

-- Flow Measurement --

 

 

Flow lower limit:

400 ml/min

Flow upper limit:

1500 ml/min

Set the upper and lower alarm limits for sample flow as desired.

The actual flow rate displays below the line.

Sample flow:

956 ml/min

Measure

Back…

Figure 3-22. Flow Measurement

3-22 Operation