Emerson NGA 2000 TO2 Section Maintenance and Service, Overview Electrical Shock Hazard

Models: NGA 2000 TO2

1 78
Download 78 pages 47.7 Kb
Page 61
Image 61

Model NGA 2000 TO2

Instruction Manual

760008-A September 2001

SECTION 4

MAINTENANCE AND SERVICE

4-1 OVERVIEW

DANGER.

ELECTRICAL SHOCK HAZARD

Operate this equipment only when covers are secured. Servicing requires access to live parts which can cause death or seri- ous injury. Refer servicing to qualified personnel. For safety and proper perform- ance, this module must be connected to a properly grounded three-wire source of electrical power.

DANGER.

POSSIBLE EXPLOSION HAZARD

This equipment is not designed and should not be used in the analysis of flammable samples. Use of this equipment in this way could result in explosion and death.

The TO2 Analyzer Module requires very little maintenance during normal operation.

The sensor in the TO2 utilizes a liquid elec- trolyte. When measuring dry gases, it may be necessary to replenish the liquid by adding distilled or deionized water.

The sensor is designed to hold at least 100 cc of electrolyte. Typically, bone dry sample gas can extract approximately 5-10 cc of water per month from the sensor. It is recommended to check the electrolyte level every 3-4 months to assure that the electrolyte level is within the acceptable operating limits as indicated by the label on the reservoir section of the sensor.

The TO2 analyzer module is equipped with a low electrolyte alarm which indicates when replenishment of the sensor is required. Please refer to the Platform manual for details on configuring alarms.

WARNING.

HIGH PRESSURE GAS CYLINDERS

This analyzer requires use of pressurized gas. See General Precautions for Handling and Storing High Pressure Cylinders page P-3.

WARNING.

PARTS INTEGRITY

Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory docu- mented components for repair.

4-2 SENSOR

CAUTION.

CAUSTIC LIQUID

Electrolyte is a caustic solution. Review the Material Safety Data Sheet in the rear of this manual.

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-1

Page 61
Image 61
Emerson NGA 2000 TO2 Section Maintenance and Service, Overview Electrical Shock Hazard, Sensor Caustic Liquid