Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

Span Calibration

Depending upon the nature of your appli- cation, it may be beneficial to verify the

span calibration of the analyzer module every 3-4 months. The following proce- dure illustrates how to initiate a span cali- bration.

Main Menu

Analyzer and I/O, expert controls and setup…

Analyzer module controls… (Or Analyzer module setup…)

TO2

37.50 ppm

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

Status…

Back…

Introduce a suitable span gas to the analyzer and allow the reading to stabilize.

Move the cursor to the “Span gas:” line and press the key. Enter the correct span gas value by using the and keys to adjust the value and the and keys to select the digit. Press the key again to save the value.

TO2

37.50 ppm

-- Calibration --

Gas scale factor…

Reset calibration!

Calibrate to span gas!

Span gas:

90.0 ppm

Measure

Status…

Back…

Move the cursor to the “Calibrate to span gas!” line and press the key. The new span calibration will be calculated.

Figure 3-17. Span Calibration

3-16 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 40
Image 40
Emerson NGA 2000 TO2 instruction manual Span Calibration, Span gas 90.0 ppm Measure Status… Back…