Model NGA 2000 TO2

Instruction Manual

760008-A September 2001

Main Menu

Analyzer and I/O, expert controls and setup…

Analyzer module controls… (Or Analyzer module setup…)

TO2

37.50 ppm

 

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

Status…

Back…

Move the cursor to the “Calibration…” line and press the key. The “Calibration” menu will display as shown below.

TO2

 

37.50 ppm

 

 

 

 

 

 

-- Calibration --

 

 

 

 

 

 

 

 

 

 

Gas scale factor…

 

 

 

 

 

Reset calibration!

 

 

 

 

 

Calibrate to span gas!

 

 

 

 

 

Span gas:

 

 

90.0 ppm

 

 

 

 

 

 

 

 

 

 

 

Measure

 

Status…

 

 

Back…

 

 

Move the cursor to the “Gas scale factor…” line and press the key. The “Gas Scale Factor 1/0” menu will display as shown on the next page.

 

TO2

-- Gas Scale Factor (1/2) --

37.50 ppm

 

 

 

 

 

 

Air background gas concentration:

 

0.0%

 

 

C2H4 background gas concentration:

 

0.0%

 

 

CO background gas concentration:

 

0.0%

 

 

CH4 background gas concentration:

 

0.0%

 

 

N2 background gas concentration:

 

100%

 

 

He background gas concentration:

 

0.0%

 

 

H2 background gas concentration:

 

0.0%

 

 

NH3 background gas concentration:

 

0.0%

 

 

C2H6 background gas concentration:

 

0.0%

 

 

C3H6 background gas concentration:

 

0.0%

 

 

Measure

 

 

Back…

 

 

Enter the volumetric percentages of each component present in the sample gas.

Move the cursor to the desired line and press the key. Then press the and keys to adjust the value and the and keys to select the digit. Press the key again to save the value.

TO2

 

37.50 ppm

 

 

 

-- Gas Scale Factor (2/2) --

 

 

C4H10 background gas concentration:

0.0%

 

 

C6H14 background gas concentration:

0.0%

 

 

Other background gas concentration:

0.0%

 

 

Other background gas factor:

100.0%

 

 

Compute new adjusted gas scale factor!

 

 

 

 

 

 

Adjusted gas scale factor:

1.00000

 

 

Scale factor computation result:

 

OK

 

 

Measure

 

 

Back…

 

 

To view and adjust additional background gas components, press the F5 key to display the second menu.

After all the background gas concentrations have been entered, move the cursor to the “Compute new adjusted gas scale factor!” command and press the key. The analyzer will automatically calculate the new gas scale factor and display the result.

Figure 3-16. Calibration – Gas Scale Factor

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-15

Page 39
Image 39