Emerson NGA 2000 TO2 instruction manual Communication protocol, Relay Outputs Setup, Invert signal

Models: NGA 2000 TO2

1 78
Download 78 pages 47.7 Kb
Page 54
Image 54

Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

Communication protocol:

AK

MODBUS RTU

None (not applicable to TO2)

NOTE:

The “special protocol definitions…” line accesses a submenu for setting the parameters of the AK and MODBUS TRU communication protocols.

TO2

37.50 ppm

-- AK Protocol Definitions --

 

 

 

Device address (RS-485 only):

1

Measure

 

Back…

 

The value can range from 1 to 50.

d.Relay Outputs Setup

From menu System SIO Module (Figure 3-25), select Relay outputs setup…

TO2

 

37.50 ppm

 

 

-- Relay Outputs Setup --

 

 

 

 

 

Output number:

1

 

 

Invert signal:

 

Disabled

 

Choose source module…

 

 

 

 

Choose signal…

 

 

 

 

Signal comes from:

Control Module

 

Signal name:

 

Failure

 

Actual status:

 

Off

 

Measure

 

Back…

 

 

open) or NC (normally closed). Full details of the SIO board are contained in its own manual.

The three lines displayed at the bottom of the “Relay Outputs Setup” menu show the current status of the selected relay output.

Signal comes from: The module chosen from the “Choose Source Module” menu.

Signal name: The signal chosen from the “Choose Signal” menu.

Actual status: The current status of the singal; Off or On.

Output number:

Corresponds to the relay number 1-3.

Invert signal:

“Disabled” signal is normal, “Enabled” signal is inverted.

Choose source module..

Brings up menu shown below.

TO2

37.50 ppm

 

-- Choose Source Module --

Control module: 0.0

TO2: 1.0

Measure

<<<

Back…

>>>

Choose desired source module for the relay output number (1-3) being config- ured.

Select the “Configuration of relay out- puts…” line in the “Local SIO Configura- tion Parameters” menu to change to the submenu “Relay Output Setup” to attach signals to the relay outputs.

There are three relays on the SIO board. The contact logic can be set with a jumper on the SIO board to select NO (normally

The list of modules will depend on the in- stalled modules.

3-30 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Page 54
Image 54