a.Menu Lines & Softkey Functionality

Menu lines can be selected with the key or the key. The selected line is dis- played as white lettering on a black back- ground (reverse text). Menus can contain four different types of lines:

Menu Line – A line ending with three dots

(…) indicates that it leads to a submenu. The submenu can be activated by press- ing the key or the key when the line is highlighted.

Function Line – A line ending with an exclamation point (!) indicates that it will start a function. The function can be acti- vated by pressing the key or the key when the line is highlighted.

Variable Line – A line ending with a colon

(:)indicates that it displays a module variable parameter. Some parameters can be changed and some parameters display only a status and cannot be changed. Paramters that cannot be changed will be displayed below a line within the menu.

Text Line – A line without punctuation marks displays information.

Tag Line – At the top of each menu screen is the tag line of the current chan- nel. To the right of the Tag is the value of the indicated channel.

The Function Keys (Softkeys) can some- times be assigned as Functions (excla- mation point) or Submenus (three dots) as shown in Figure 3-2.

Tag Line

Menu Line

Function Lines

Variable Lines

CLD

7.50 ppm

 

-- Main Menu --

Analyzer basic controls (calibration) & setup…

Analyzer and I/O, expert controls & setup…

System configuration and diagnostics…

Display controls…

Time & Date:

System tag:

Measure

 

Status…

 

Channel

 

Lock…

 

MFG Data

 

 

 

 

 

 

 

 

 

 

 

Selected Line

(Reverse Text)

Lines below this separator are information and cannot be changed.

Function Keys F1 – F5 Legend

Figure 3-2. The Display Screen

b.Common Function Keys

The function keys are shown in Figure 3-1 And Figure 3-2.

Display – Change from the single com- ponent display to the multi-component display. F1 in the single component dis- play.

Measure – Change from menus and submenus to the single component dis- play of the selected channel. F1.

Status – Change to the menu “Current measurement parameters” which displays the most important parameters and infor-

mation about the status of the current channel or module. F2 if available. (See Section 3-4 )

Main – Change from single component display to the main menu. F3 in the single component display. (See Section 3-2e)

HOME – Change for various menus to the main menu. F1.

Channel – Scrolls through the channels in the same menu. In the main menu and the single component display menu it moves between the channels of the con- nected Analyzers and Analyzer modules. In the submenus it moves only between