Model NGA 2000 TO2
Instruction Manual
Sensor model:
Adjustable for both “Standard” and “Sta- bel” models in the following ranges:
Clamp low PVA to 0.02:
This parameter is used for very low oxy- gen concentration levels where the sen- sor current may be unstable about zero. Normally this would be set to “No.”
ppm to mg/Nm3 conversion factor:
The ppm to mg/Nm3 conversion factor in the range of 1 to 100000. The conversion factor depends on the sample gas used.
Lower [Upper] explosion limit (LEL) [UEL]:
The parameters are used for the CLD and PMD modules and do not apply to the TO2 They should both be set at 0.00%.
Temperature compensation at span gas:
The “Stabel” sensor model contains a special electrolyte that is less susceptible to small concentrations of acids in the sample stream.
This parameter is used to set the upper temperature limit for the sensor which can occur at the upper range of oxygen con- centration. The range is
Rosemount Analytical Inc. A Division of Emerson Process Management | Operation |