Emerson NGA 2000 TO2 Linearization…, 0000 ppm, Linearization raw values sensor current…, 3828 uA

Models: NGA 2000 TO2

1 78
Download 78 pages 47.7 Kb
Page 65
Image 65

Model NGA 2000 TO2

Instruction Manual

760008-A September 2001

TO2

37.50 ppm

 

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

 

Status…

 

Back…

 

TO2

Y1:

Y2:

Y3:

Y4:

Measure

37.50 ppm

-- Linearization Set Point Values --

0.0000 ppm

1.050 ppm

7.000 ppm

72.800 ppm

Back…

Select Linearization…

TO2

37.50 ppm

 

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

 

Back…

 

Select Linearization raw values (sensor cur- rent)…

TO2

37.50 ppm

 

-- Linearization Raw Values --

 

 

X1:

0.3828 uA

X2:

3.0400 uA

X3:

20.9300 uA

X4:

193.000 uA

Measure

 

Back…

 

Enter the data for each raw value point.

Return to the previous menu (Back…)

TO2

37.50 ppm

 

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

 

Back…

 

Select Linearization set point values (ppm values)…

Enter the data for each set point value. Return to the previous menu (Back…)

TO2

37.50 ppm

 

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

 

Back…

 

Select Linearization temperature values…

TO2

 

 

37.50 ppm

 

 

 

-- Linearization Temperature Values --

 

 

 

 

 

 

 

 

T1:

 

 

25.0 °C

 

 

T2:

 

 

25.0 °C

 

 

T3:

 

 

25.0 °C

 

 

T4:

 

 

25.0 °C

 

 

 

 

 

 

 

 

Measure

 

Back…

 

 

Enter the data for each temperature value point..

Return the the previous menu (Back…)

TO2

37.50 ppm

 

-- View & Modify Linearization Table --

Linearization raw values (sensor current)…

Linearization set point values (ppm values)…

Linearization temperature values…

Load linearization data!

Measure

 

Back…

 

Select Load linearization data! press enter. The new data is now loaded into the analyzer and accurate measurements can be obtained.

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service

4-5

Page 65
Image 65
Emerson NGA 2000 TO2 instruction manual Linearization…, 0000 ppm, Linearization raw values sensor current…, 3828 uA, 25.0 C