HP 16500C, 16501A LOGIC manual Measurement Initialization Error

Models: 16500C 16501A LOGIC

1 130
Download 130 pages 6.88 Kb
Page 109
Image 109

If You Have a Problem

“Measurement Initialization Error”

“Measurement Initialization Error”

This error occurs when you have installed the cables incorrectly for one or two HP 16550A logic analysis cards. The following diagrams show the correct cable connections for one-card and two-card installations. Ensure that your cable connections match the drawing, then repeat the measurement.

Cable Connections for One-Card HP 16550A Installations

 

Cable Connections for Two-Card HP 16550A Installations

See Also

The HP 16550A 100-MHz State/500-MHz Timing Logic Analyzer Service

 

Guide.

5–13

Page 109
Image 109
HP 16500C, 16501A LOGIC manual Measurement Initialization Error